Re: So I got myself a Fluke 9010a and...

From: John Robertson <jrr_at_flippers.com>
Date: Thu Apr 10 2003 - 01:09:06 EDT

<x-flowed>
This is the RESET showing up...find the Watchdog Disable test point and
connect it to ground. Then try your tests again.

This is covered in the 9010A Operators manual, copies are available in PDF
if you don't have a hard copy...

John :-#)#

At 07:48 PM 09/04/2003 -0500, Joe wrote:

> So I got myself a Fluke 9010a and I power it up, pod passes self test
>just fine.. I plug the beast into the Gravitar CPU socket, all good so
>far. Power up the 9010a and then the Gravitar. Now I do a RAM SHORT test
>from 0000-07FF and get the following error "Active Force Line @
>0000-Loop?"
>
>So, what am I doing wrong? The Gravitar board I'm using is a known good
>board, so this must be operator error, right?
>
>Any help or suggestions will be appreciated.
>
>Thanks,
>JB