Re: sencore cr70 repair

From: Darren Finck <dgfinck_at_yahoo.com>
Date: Wed Feb 27 2013 - 22:27:02 EST

The problem is that after replacing IC4 it's no longer calibrated and it's failing the test procedure. The voltage on G1/G2 is supposed to swing from 20 to 400 volts but it's actually swinging from 80 to about 460 volts

Can anybody help explain the purpose of pots R24, R23, and how I might want to adjust them to recalibrate the unit properly? It looks like I need to adjust both of these but I don't know how.
-joey

On Tue, Jan 15, 2013 at 10:45 AM, Joey Oravec <joeyoravec@gmail.com> wrote:

Have you scanned the schematics?  If so, send a copy (or a link to them) and I'll be glad to look at them and try explaining the purpose of the pots you're asking about.  I have a CR31, and had to dig pretty deep into it to repair it, so I became pretty familiar with its schematics & operation.  So with the expectation that the CR70 is similar to the CR31, I should be able to follow it pretty well.

---------------------------------------------------------------------------