Re: Fluke 9010A Input voltage

From: John Robertson <jrr_at_flippers.com>
Date: Mon Oct 31 2005 - 16:14:44 EST

Good point Rodger, I think that was hidden in the yada, yada part of
the instructions for changing voltage that I didn't quote...

Now to find the PDF of the 9010A manuals. They aren't up on the ftp site:
ftp://ftp.flippers.com/TTL/Test%20Equipment/ the backup CD is buried
somewhere in my shop...

John :-#)#

At 3:40 PM -0600 10/31/05, Rodger Boots wrote:
>Be careful, a lot of times when you change voltage you also should
>change the fuse.
>In the case of the 9010A the line fuse should be 1A fast blow for
>120 V and 1/2 A fast blow for 240 V.
>
>
>
>Filip Hautekeete wrote:
>
>>Hello Steve,
>> You are correct. Thanks to John Robertson I figured out there is a
>>slide card for voltage selection behind the fuse panel. No need to
>>drag the systems out of your garage ;-)
>> Thanks for your help!
>> Filip